PROMPTU Pressura Transducer
Transducis pressio disponibilis est ad continuam mensuram pressionis physiologicae et determinationis aliorum parametri haemodynamicae magni momenti.Hisern's DPT praebere potest mensuras accurate et certas sanguinis pressionis arteriarum et venarum in operationibus cardiacis intervenientibus.
Indicatur pressionis vigilantia applicationes sicut:
Pressura sanguinis arteriae (ABP)
Venarum vis centralis (CVP)
Intra cranii pressura (ICP)
Intra abdominis pressura (IAP)
Flushinga Fabrica
Valvae Micro-porosa rutilantia, in rate fluente constanti rutilantia, coagulationem in pipelino vitandam et corruptelam waveform ne
Duae influunt rates 3ml/h et 30ml/h (pro neonatis) ambo praesto sunt
Lavari potest, tollendo et trahendo, facile ad operandum
Special tribus-modo Stopcock
Flexibile transitum, commodum rutilant et exinanitionem
Praesto cum systemate sampling sanguinis clauso, reducendo in periculo infectionum nosocomialium
Automatic rutilant ne coagulationis et bacterial coloniatio
Complete Specifications
Varia exempla diversis necessitatibus occurrere possunt, ut ABP, CVP, PCWP, PA, RA, LA, ICP, etc.
VI genera connexionum componi possunt cum notis maxime monitorum in mundo
Inscriptiones multi-colores, monitiones ad monitor sanguinem pressurae manifestae
Praebere album non-porosum pileum reponere ad vitare infectio nosocomial
Libitum sensorem possessor, plures transductores figere potest.
Funem adaptatorem libitum, compatible cum monitores variarum notarum
ICU
Operans volutpat
Urgentis ātrium medicīnae
Cardiology Department
Anesthesiology Department
Interventus Therapy Department
ITEMS | MIN | TYP | MAX | UNITS | NOTAE | |
Electrical | Operans Impetus dolor | -50 | 300 | mmHg | ||
Per impressionem | 125 | psi | ||||
Nulla Pressura cinguli | -20 | 20 | mmHg | |||
Input Impedimentum | 1200 | 3200 | ||||
Impedimentum output | 285 | 315 | ||||
Output Symmetria | 0.95 | 1.05 | Ratio | 3 | ||
Copia voltage | 2 | 6 | 10 | Vdc or Vac rms | ||
Periculum Current (@ 120 Vac rms, 60Hz) | 2 | uA | ||||
Sensibilitas | 4.95 | 5.00 | 5.05 | uU/V/mmHg | ||
Euismod | Calibration | 97.5 | 100 | 102.5 | mmHg | 1 |
Linearitas et Hysteresis (-30 ad 100 mmHg) | -1 | 1 | mmHg | 2 | ||
Linearitas et Hysteresis (100 ad 200 mmHg) | -1 | 1 | % output | 2 | ||
Linearitas et Hysteresis (200 ad 300 mmHg) | -1.5 | 1.5 | % output | 2 | ||
Frequentia Responsio | 1200 | Hz | ||||
Drift cinguli | 2 | mmHg | 4 | |||
Scelerisque Span Shift | -0.1 | 0.1 | %/°C | 5 | ||
Shift cinguli scelerisque | -0.3 | 0.3 | mmHg/°C | 5 | ||
Status Shift (@ 5KHz) | 5 | Gradus | ||||
Defibrillator sustinere (CD joules) | 5 | obit | 6 | |||
Lux Sensibilitas (3000 Foot Candle) | 1 | mmHg | ||||
Enviromental | Sterilisatio (ETO) | 3 | Cycles | 7 | ||
Operating Temperature | 10 | 40 | °C | |||
Repono Temperature | -25 | +70 | °C | |||
Operating Product Vita | 168 | Horae | ||||
Fasciae vita | 5 | Annos | ||||
Dieelectric Naufragii | 10,000 | Vdc | ||||
Umor (externi) | 10-90% (non-condensatione) | |||||
Media Interface | Dielectric Gel | |||||
Fermentum Tempus | 5 | Secundis |